Perform Verification of Operations Test on the Odyssey® M Imager
For questions about performing the Basic Operations Test, please contact your LI‑COR representative.
-
Clean the scanning surface with 70% ethanol and lint free wipes. The scanning surface must be free of lint and completely dry before the test pattern is placed on the surface.
-
Place the test target 6533-200 (from packet 9933-105) face down on the scanning surface.
- Start LI‑COR Acquisition Software
, if you have not already.
- Select your Username and Image Data Folder.
- Click Scan
.
- Choose the imager you want to connect to.
- Click Connect.
- Click the imager name displayed in the top left corner of the LI‑COR Acquisition Software window.
- In the menu that opens, click Perform BOT.
- Click .
- Once the scans have completed, click .
- Choose where to export the file.
- Email the exported file with the Support Scientist who shared this document with you.
- In the email, use the subject line
BOT ODM-XXXX
. - The LI‑COR Service Team will review the information and verify the system is working properly.