Perform Verification of Operations Test on Odyssey® XF or Odyssey Fc
-
Remove the Odyssey Fc test target from its storage sleeve and place it in a well-lit area for a minimum of 3 minutes to charge the phosphorescent portion.
The target used for this test must be in excellent condition. Do not use any target that has not been stored properly or that has been contaminated or damaged in any way.
-
Push the drawer button and ensure that the drawer opens all the way.
-
Place the test target in the drawer by itself with the printed side up. Do not use a tray.
-
Push the drawer button again. Ensure the drawer closes all the way without making any unusual sounds that may indicate that the motor is running against the mechanical stop.
- Start LI‑COR® Acquisition Software
, if you have not already. - Select your Username and Image Data Folder.
- Click Scan
.
- Choose the imager you want to connect to.
- Click Connect.
- Click the imager name displayed in the top left corner of the LI‑COR Acquisition Software window.
- In the menu that opens, select .
- Click .
- Once the scan is complete, click .
- Choose where to export the file. The exported file will have a
.lipoextension. - Email the exported file to the Support Scientist who shared this document with you.
- In the email, use the subject line
VNOT OXF-XXXX. - The LI‑COR Service Team will review the information and verify the system is working properly.