Welcome to the LI‑COR® Acquisition Software Help!
Scan Multiwell Plates on the Odyssey® M Imager
Once connected, click Multiwell Plate on the Choose Assay page to begin imaging a plate-based assay.
Choose Plate Assay
Choose the type of plate assay you want to image.
-
In‑Cell Western - Use this workflow to acquire In‑Cell Western or On-Cell Western plate images.
Both the In‑Cell Western and On-Cell Western are quantitative immunofluorescence assays performed in microwell plates. Data analysis is based on whole-well fluorescence quantification.
Available image settings for In-Cell WesternThe following image settings are available for the In-Cell Western option.
-
Resolution for 96-Well plates: 100 µm
-
Resolution for 384-Well plates: 50 µm
-
Focus Offset: 3.8
-
Available Focus Offset Range: 1.00 - 5.00 mm
-
-
Absorbance Assay
Available image settings for Absorbance Assay-
Resolution for 96-Well Plates: 100 µm
-
Resolution for 384-Well Plates: 50 µm
-
Focus Offset: 3.00 mm
-
Available Focus Offset Range: 1.00 - 5.00 mm
-
-
Cell Analysis - Cell Analysis refers to a broad range of assays for studying the function and localization of proteins in living and fixed cells. Cell Analysis covers assays that are not more precisely described as one of the other Multiwell Plate Assay types.
Available image settings for Cell AnalysisThe following image settings are available for the Cell Analysis option.
-
Resolution: 100 µm (96-well, 48-well, 24-well, 12-well, 6-well)
-
Focus Offset: 3.00 mm
-
Available Focus Offsets: 1.00 - 5.00 mm
-
Select plate locations
Select the location of your plate(s) on the Odyssey M Scan Surface.
Choose detection options
-
If you have multiple Scan Areas, click Edit Channels to choose detection options independently for each plate.
-
Select the plate's well format.
-
Click Scan to begin image acquisition.
Detection label reference
The detection labels listed in this table are examples. If you do not see the detection label you want to use, check its spectral characteristics to make sure it is compatible.
|
Channel |
Labels |
Excitation |
Emission |
|---|---|---|---|
| 800 | IRDye® 800CW, DyLight 800 | 785 nm | 816 - 840 nm |
| 700 | IRDye 680RD, CellTag™ 700 Stain, Sapphire700™ Stain, DRAQ5 | 685 nm | 721 - 740 nm |
| 520 | VRDye 549, Cy3, Alexa Fluor 546, CellTag 520 Stain | 520 nm | 570 - 610 nm |
| 488 | VRDye 490, FITC, Alexa Fluor 488 | 488 nm | 519 - 543 nm |
| Chemi | HRP and AP substrates, luminescent assays | NA | NA |
Any combination of up to four of the channels in the table can be selected per plate per acquisition. Different channels can be selected for different plates.
| Channel | Labels | Excitation | Emission |
|---|---|---|---|
| 630 Trans | TMB - non stopped, Bradford Protein Assay, Pierce 660nm Protein Assay, Modified Lowry Protein Assay | 631 - 659 nm | Clear Window |
| 525 Trans | OPD, BCA Protein Assay | 500 - 530 nm | Clear Window |
| 470 Trans | OPD, TMB - stopped, BCA Protein Assay | 447 - 472 nm | Clear Window |
| RGB Trans | True color | Transmissive RGB Color | |
| Channel | Labels | Excitation | Emission |
|---|---|---|---|
| 800 | IRDye® 800CW, DyLight 800 | 785 nm | 816 - 840 nm |
| 700 | IRDye 680RD, CellTag™ 700 Stain, Sapphire700™ Stain, DRAQ5, SYTO 60 | 685 nm | 721 - 740 nm |
| 520 | VRDye 549, Cy3, Alexa Fluor 546, Red fluorescent protein (RFP), Propidium Iodide | 520 nm | 570 - 610 nm |
| 488 | VRDye 490, Green Fluorescent Protein (GFP), SYTOX Green | 488 nm | 519 - 543 nm |
| Chemi | HRP and AP substrates, luminescent assays | NA | NA |
Any combination of up to four of the channels in the table can be selected per plate per acquisition. Different channels can be selected for different plates.
Scan
After clicking Scan on the Choose detection options page, the Scan page will open with the image acquisition process already started.
When the scan has completed, the Ready symbol will appear.
Stop Scan
If you need to stop an image acquisition before the acquisition process completes, click .
This will stop all pending Acquisitions and will remove any partially completed Acquisitions.
Export Acquisition
Ensure the Include all acquisitions option is selected to export every image for every channel from this acquisition process.
Click to export images to a single file. Import the file into Empiria Studio® Software for analysis.
Click to export image file(s). Images exported using this method are not intended for analysis.
Images can be exported as high resolution TIFF files (for journal publications) or as PNG files appropriate for digital presentation (such as a slide show).
When you are exporting a single image, the default export image size shown in the "Export" dialog is the largest dimension (i.e., width or height) of the image set to 10 cm (4 in), and the other dimension is set according to the aspect ratio of the image. If you change the width or height in the dialog, the other dimension will update automatically to maintain the correct aspect ratio.
See View Images in the Image Gallery for more information.
Workflow
- Click to return to the beginning of the workflow to acquire another multiwell plate image.
- Click Rescan to acquire new image(s) with the same settings.
- Click Home to disconnect from the imager and return to the Home page.