Scan a Gel on the Odyssey® M Imager
Once connected, click Gel on the Choose Assay page to begin imaging gels.
The following imaging settings are used in this workflow. Other settings can be entered in the Custom workflow.
-
Resolution: 100 µm
-
Default Focus Offset: 0.50 mm
Set the Focus Offset to approximately half the thickness of the gel. The default Focus Offset is 0.50 mm (approximately half the thickness of a typical precast protein gel).
Set Scan Area
- Draw multiple Scan Areas to image multiple gels.
- Adjust the exact size and location of a Scan Area. Select a Scan Area, then adjust the number in the X location, Y location, Width, and Height fields in the Control Panel.
Choose detection options
If you have multiple Scan Areas, click Edit Channels to choose detection options independently for each Scan Area.
Click Scan to begin image acquisition.
-
A single Fluorescent, non-RGB Trans or RGB Trans channel may be selected.
-
The 700 channel should be used if quantification of Coomassie stained protein gels is desired.
-
For basic gel documentation of Coomassie gels where quantification is not required the 630 Trans or RGB Trans channel may be preferred for acquisition speed (about 30 seconds for a 10 x 10 cm gel vs. 2 minutes for the 700 channel).
Detection label reference
The detection labels listed in this table are examples. If you do not see the detection label you want to use, check its spectral characteristics to make sure it is compatible.
-
Non-RGB trans channels cannot be imaged together.
-
Non-RGB trans channels cannot be imaged with a fluorescent channel.
-
The RGB trans channel can be imaged with a non-RGB trans channel.
-
The RGB trans channel can be imaged with one or more fluorescent channels.
-
Up to three fluorescent channels can be imaged together.
| Channel | Labels | Excitation | Emission |
|---|---|---|---|
| 800 | IRDye® 800CW, DyLight 800 | 785 nm | 816 - 840 nm |
| 700 | IRDye 680LT, IRDye 680RD, SYTO 60, Coomassie Stain | 685 nm | 721 - 740 nm |
| 520 | GelRed, Ethidium Bromide, Midori Green, Pro-Q Diamond Phosphoprotein Gel Stain | 520 nm | 570 - 610 nm |
| 488 | SYBR Safe, SYBR Green, GelGreen, GreenView, Pro-Q Emerald 488 Glycoprotein Gel Stain | 488 nm | 519 - 543 nm |
| 488A | SYPRO Ruby, SYPRO Orange, SYBR Gold | 488 nm | 570 - 610 nm |
| RGB Trans | Colorimetric Protein MW Markers | NA | NA |
| 630 Trans | Coomassie Stain | 631 - 659 nm | Clear Window |
| 470 Trans | Silver Stain | 447 - 472 nm | Clear Window |
Scan
After clicking Scan on the Choose detection options page, the Scan page will open with the image acquisition process already started.
When the scan has completed, the Ready symbol will appear.
Stop Scan
If you need to stop an image acquisition before the acquisition process completes, click .
This will stop all pending Acquisitions and will remove any partially completed Acquisitions.
Export Acquisition
Ensure the Include all acquisitions option is selected to export every image for every channel from this acquisition process.
Click to export images to a single file. Import the file into Empiria Studio® Software for analysis.
Click to export image file(s). Images exported using this method are not intended for analysis.
Images can be exported as high resolution TIFF files (for journal publications) or as PNG files appropriate for digital presentation (such as a slide show).
When you are exporting a single image, the default export image size shown in the "Export" dialog is the largest dimension (i.e., width or height) of the image set to 10 cm (4 in), and the other dimension is set according to the aspect ratio of the image. If you change the width or height in the dialog, the other dimension will update automatically to maintain the correct aspect ratio.
See View Images in the Image Gallery for more information.
Workflow
- Click to return to the beginning of the workflow to acquire another gel image.
- Click Rescan to acquire new image(s) with the same settings.
- Click Home to disconnect from the imager and return to the Home page.