Custom Scan for the Odyssey XF Imager or Odyssey Fc Imager
The following settings are available when using the Custom workflow.
-
Resolution Options: 125 µm
-
Acquire times:
Choose detection options
The detection labels listed in this table are examples. If you do not see the detection label you want to use, check its spectral characteristics to make sure it is compatible.
| Channel | Labels | Excitation | Emission | Combined Selections Allowed |
|---|---|---|---|---|
| 800 | IRDye® 800CW, DyLight 800 | 785 nm | 816 - 846 nm | 600, 700, Chemi |
| 700 | IRDye 680LT, IRDye 680RD, Revert 700 Total Protein Stain, Alexa Fluor 680 | 685 nm | 716 - 746 nm | 600, 800, Chemi |
| Chemi | HRP and AP chemiluminescent substrate | NA | NA | 600, 700, 800 |
| 600 | SYBR Safe, Ethidium Bromide | 520 nm | 573 - 637 nm | 700, 800, Chemi |